Effective August 2, 2021, Clinical Pathology Laboratories is pleased to announce a change in instrumentation/ methodology and reference range for Serum Myoglobin with transition to the Roche Cobas electrochemiluminescence immunoassay (ECLIA) method. Previously, Serum Myoglobin testing was performed on the Beckman Coulter DxI by paramagnetic particle chemiluminescent immunoassay.

The determination of myoglobin in serum is an important factor in the diagnosis of acute myocardial infarction (AMI), early reinfarction and successful reperfusion following lysis therapy.

Changes in reporting:

Serum Myoglobin 4249

Male: 17 -106 ng/dL

Female: 14 – 66 ng/dL

Male: 28-72 ng/dL

Female: 25-58 ng/dL

Assay Test Code Prior Reference Range New Reference Range

 

Order Unit Codes and Test Names: 4249 Myoglobin, Serum
Specimen Requirements:

1 mL serum.  Allow sst to clot in an upright position for at least 30 minutes, then centrifuge sample within 2 hours of collection. Transfer serum to a plastic transport tube. Refrigerate. 

Do not collect samples from patients receiving therapy with high biotin doses (>5 mg/day) until at least 8 hours following the last biotin administration.

Transport Temperature: Refrigerated
Stability (collection to initiation of testing):

Room Temperature 15-25°C: 7 days

Refrigerated 2-8°C: 7 days

Frozen (-15) - (-25)°C: 1 month

Performed: Monday through Friday / PM Shift
Analytic Time: 1 day
CPT Code: 83874